Endoscopes still contaminated after cleaning, study shows

Published: 6-Aug-2015

Colonoscopes and gastroscopes can harbour viable microbes, even when recommended reprocessing guidelines are adhered to


Potentially harmful bacteria can survive on endoscopes used to examine the interior of the digestive tract, despite a multi-step cleaning and disinfecting process, according to a new study published in the American Journal of Infection Control, the official publication of the Association for Professionals in Infection Control and Epidemiology (APIC).

Though endoscopes were cleaned in accordance with guidelines, microbes and residual contamination remained on surfaces after each stage of cleaning, according to study findings.

Researchers from Ofstead & Associates in Saint Paul, Minnesota, US and Mayo Clinic in Rochester tested samples collected from 60 encounters with 15 colonoscopes and gastroscopes used for gastrointestinal procedures after each reprocessing step to assess contamination levels. They observed all reprocessing activities, using a checklist to ensure that cleaning protocols were performed in accordance with published guidelines.

This study demonstrates that colonoscopes and gastroscopes can harbour residual organic material

Reprocessing consisted of: bedside cleaning, manual cleaning in dedicated reprocessing rooms, and automated endoscope reprocessing with a high-level disinfectant. Disinfected endoscopes were stored vertically after drying with isopropyl alcohol and forced air. When contamination levels exceeded pre-determined benchmarks for each cleaning step, technicians went beyond guidelines and repeated cleaning procedures, retesting after each attempt to reduce contamination.

Researchers performed microbial cultures and various rapid tests to detect viable organisms and organic residue that remained after each step of cleaning.

Viable organisms were detected on 92% of devices after bedside cleaning; 46% after manual cleaning; 64% after high-level disinfection, and 9% after overnight storage. Rapid indicator tests detected contamination above benchmarks on all devices after bedside cleaning; 92% after manual cleaning; 73% after high-level disinfection, and 82% after overnight storage.

'This study demonstrates that colonoscopes and gastroscopes can harbour residual organic material, including viable microbes, even when adherence with recommended reprocessing guidelines is verified,' said the study authors.

'More research is needed to identify processes that can ensure all flexible endoscopes are free of residual contamination and viable microbes prior to patient use, including the potential use of routine monitoring with rapid indicators and microbiologic cultures. Results from this study suggest that current standards and practices may not be sufficient for detecting and removing residual contamination.'

Potential limitations of the study include that it was only a single-site study. In addition, reprocessing technicians were aware of the researchers' use of a checklist to ensure guideline compliance and therefore may have devoted more time and effort to reprocessing.

Another caveat is that technicians were immediately informed about contamination that exceeded benchmarks and repeated cleaning steps – actions that are not generally part of standard practice.